There were 22 articles found in this category:

  1. questionThe installation of the CM/IM900 has been performed correctly without any errors and still the LED on the camera is RED
    We have seen this situation recently appear more often. If the LED of the camera is red after a succesful installation the camera informs the connection is not a full USB 2.0 speed. Check in the BIOS setting of the PC is the USB port speed is set to 2.0
  2. questionI have problems viewing the captured movie
    If the display of captured movies does not work on your PC (it seems that EyeCap freezes or an error message is displayed) please install the FFDshow_CoDec. This CoDec can be found on the Internet and on the www.eyecap.com website. Make sure that during the installation of this CoDec you select ...
  3. questionThe user want to capture images with the IM/CM 900 on EyeCap with one click and does not want to use the History Trigger function in EyeCap
    Standard EyeCap has enabled the history trigger function to select the best image after you have pressed the capture button on the trigger module.The user has to press the capture button twice: - once to freeze the image and - second to select the best image from the history trigger, or just cli ...
  4. questionThe button on the release module for the CM/IM 900 do not work
    The release module simulates the pressing on the CTRL-ALT and F6 till the F12 keys on the keyboard. If other software is installed on the PC which also uses these key combination (e.g. display drivers which used the CTRL-ALT + F-keys as hot-keys to change settings) then the release module does ...
  5. questionDo you have a manual how to use the IM900 with EyeCap
    Please find below in the attachment section the Quick User Guide for the IM900 with EyeCap
  6. questionThe video on the CM/IM900 seems to be choppy
    Most likely it is a hardware problem. Please make sure you have a powerful computer with at least 2 GB of ram and preferred a dedicated graphics card (check recommended hardware specifications). Make sure you use a good quality USB cable. Also check if all MS windows updates and the FFD-show cod ...
  7. questionWhat are the settings for the CM and IM-900
    Please download below PDF file (see below) for the settings
  8. questionWhy should I purchase the CM900 or IM900 rather then a consumer camera
    Using a consumer camera is less expensive but not necessarily a simple method of slit lamp imaging. If it was simple then we would have had no need to develop the IM or CM cameras and everyone would be using consumer cameras. The lens systems on consumer cameras make the integration with the sli ...
  9. questionWhy is back ground illumination considered essential
    There are two major reasons why imaging slit lamp should have a quality background illumination system: - Overview images: for this the diffuser will work but using the background illumination and slit beam (diffused) will produce a shadow free illumination with natural colours and two light ref ...
  10. questionThe BQ with IM 900 seems very long
    The length of the system can be reduced slightly by using the CM 900. The CM900 with a Beamsplitter (44mm) is 24mm shorter than IM900 (68mm).
  11. questionWhat is the maximum lenght of an usb cable
    5 meters or with a powered hub longer : http://www.usb.org/about/faq/ans5. Make sure you use good quality cables.
  12. questionWhere can I learn about slit lamp imaging
    Courses on slit lamp imaging are run regularly by the OPS (www.opsweb.org) and OIA (www.oia.org.uk/ ) and Haag-Streit, through the local distributor, may also be able to help improve skills. Slit Lamp: Examination and Photography is recently published book by Csaba Martonyi and is an excellent r ...
  13. questionDoes the IM/CM 900 require 2GB memory
    We recommend that PC running IM and CM should have a minimum 2GB RAM. If only 1GB is available then the PC should have a separate graphics card. (i.e. Cannot share graphics with RAM).
  14. questionCan I take fundus images on the IM900
    Yes, good quality fundus images are possible and are made easier to capture due to the history trigger. Much better than Firewire camera because of the improved dynamics of the IM900 sensor.
  15. questionWhy a 2 Megapixel sensor was selected
    The movie sequence from IM 900 is full resolution of 1600 x 1200. Most other systems reduce the resolution on the sensor and compress the movie. The benefit with IM 900 is that when the doctor wants to freeze the movie (which he frequently does) it is a full still image that is displayed rather ...
  16. questionWhat is the file size of 1 second video
    With 12.5 frames per second the file size of a video of 1 second is 67.5MB (uncompressed)
  17. questionWhat is the file size of the IM900 images
    The file size for BMP images is roughly 5.4 MB. The files size of JPEG images depends on the applied compression ratio and complexity of the image.
  18. questionWhy should I use an eyepiece with the cross hairs?
    For normal use the eyepieces of all slit lamps should be set with the test rod supplied to compensate for your accommodation. Each user should do this and they may also have to take into account that their eyes may change during the course of a working day.
  19. questionCan I leave the power on the IM900 on all the time or should I switch it off. i.e. unit used on a unit. (Power from unit or power in correlation with slit lamp)
    IM900 should be powered by the unit and not switch on and off as this could present problems with the PC connection. We have seen that in some cases the RM-01 of the IM900 needs to be powered before the PC is switched on. The camera will only be on when the PC is on, despite the power from the unit.
  20. questionWhat is the field of view
    The camera captures approximately the same view as seen through the slit lamp. To be more specific, the horizontal view is close to the width of the image with a very small cut off at the top and bottom. This is because the image through the slit lamp is circular.
(Page 1 of 2)      Prev | 1 | 2 | Next