Is it possible to measure the endothelium through swollen or opaque corneas
The cornea is part of the optical system that provides the endothelial image to the specular microscope. If the cornea is compromised then this will affect the image quality. It may still be possible to image the endothelium through corneal oedema but we advise that you examine the data carefull ...
What is the accuracy of the EM935
The accuracy of the ME 935 is defined as 747Pix/mm +/- 2.5%. Best reproducibility is achieved if the endothelium images are analyzed in a semi automatic manner. This means the images are taken automatically, the area of interest is drawn only where the cells are easily visible and the automatic ...
In Alignment mode, slit does not become narrow but fades out
Check if the base plate/pin is well connected and unit is centered. If this is the case and you have still problems, please contact our service department.
Alignment mode holding hand in front of the lens image is grey
In Alignment mode the image on the screen should be black. If it is otherwise (grey) the sensitivity of the camera is not correct. Please contact our service department.
Alignment mode, with slit on 0.2 no image does appear, but on 3mm it does
Base plate height is not set correctly. Possibly the wrong base plate used, contact our service department.
Error "This unlicenced copy of IMAQ Vision may be used for demo purposes only" comes up
Please install the runtime license from CD. Look at Folder x:\Run-Time-License and start the Program vision61rte.exe
Can I use longer lengths of Firewire cables
The maximum cable length is 5m. Longer lengths will reduce the transfer speed and this will cause problems. Always use a good quality cable.
Can we place the EM935 on the left hand side
No. This is not possible because the software is programmed to search for the corneal reflex on the left of specular endothelial reflex.
What is the EM 935
EM935 is a slit lamp mounted endothelium microscope designed to make a rapid, automatic analysis of the corneal endothelium and conveniently fits onto most slit lamps. Image capture is automatic and analysis includes cell density and morphology data
Can the EM935 do any other high magnification imaging (i.e. lens pathology)
The EM935 has been designed specifically for specular reflectance images of the corneal endothelium. The magnification of ~200 x limits the use for other types of imaging. Some users have expressed an interest in high magnification black and white images and we will keep all Haag-streit distrib ...
Possible to use the EM935 for corneal grafts
The EM 935 is designed to measure the corneal endothelium ‘in vivo’ and as such has not been used on graft donor material or eyes. There are instruments available for this specialized work and we suggest the EAT from Rhine-Tec http://www.rhine-tec.de/gb-hp-produkte-EAT.htm
Can EM 935 software handle measurement of endothelium with guttata
Measuring eyes with guttata is a challenge for all specular microscopes as the software has trouble differentiating the difference between a cell and guttata. The best and fastest method is to measure automatically and then correct manually but this depends on the eye. Alternatively you can R ...
Can EM 935 software run on a touch panel pc
The processing power required by the EM935 is minimal and therefore any PC or laptop should be capable. If the Firewire port is not powered (i.e. on most laptops) then the power supply for the camera must be used. In our experience the best solution is to use a 6 pin firewire port as this means ...
Can the EM935 do any other high magnification imaging (i.e. lens pathology)
The EM935 has been designed specifically for specular reflectance images of the corneal endothelium. As it is around 200 x magnification it is not much used for anything else. In theory it could be used to image most parts of the anterior chamber at 200 x in black and white but the software is l ...